Chapter 4 Preparing LogTag(s) for use 35
ticks in both boxes next to the "After X consecutive alert readings" and "After X alert
readings have occurred".
2. You can not enter a number into the "After X consecutive alert readings" field which is
larger than the number in "After X alert readings have occurred", since the latter alert
condition would always be met first.
3. Similarly, you can not enter a number into the "After X alert readings have occurred"
field which is smaller than the number in "After X consecutive alert readings". The
software will adjust field values as appropriate.
Advanced Alert Settings
Some LogTags support more detailed alert settings than described before. If LogTag
Analyzer encounters one of these loggers, an additional
button is visible in
the bottom left of the configuration screen.
When this button is pressed, the alert configuration screen changes and displays
additional settings:
The red alert indicator can now be configured independently for upper and lower alert
values.
Function Purpose
Readings above X
Readings below Y
In addition to entering the threshold levels tick the boxes next to
"Readings above" or "Readings below" to enable the alert
conditions for only the upper or only the lower threshold
settings, or for both. If both boxes are ticked, the behaviour is
the same as for the standard alert settings. The charts will
display alert readings according to the settings made in the
Charts
section of the 'Options' Screen.